Glucometer
Glucometer
Features (Technical advantages)
- FAD-GDH system : It is not easily affected by oxygen molecule in the blood.
- User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.
Specification
| Blood volume | 0.6μL |
| Sample type | Capillary whole bloodyenous whole blood |
| Calibration | Plasma equivalent |
| Measuring time | 10s |
| Meter storage/transportation condition | -20C~55℃ |
| Dimension | 105*60*23(mm) |
| Weight | 77.2g,with battery |
| Power source | 3V DC,10mA,2 AAA alkaline batteries |
| Memory | 200 blood glucose test results |
| Operating condition | 10℃~35C≤80% RH |
| Construction | Hand-held |
| Measurement units | mg/dL or mmol/L |
| Measurement range | 20~600 mg/dL or 1.1~33.3mmo/L |
Precision: test result of Safe-Accu2 meet the requirement below:
Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:
| Concentration Range | Bias% |
|---|---|
| <5.5 mmol/L(100 mg/dL) | Within±0.83 mmol/L(15 mg/dL) |
| 25.5 mmol/L(100mg/dL) | Within ± 15% |
Precision: test result of Safe-Accu 2 meet the requirement below:
| Concentration Range | Require ment |
|---|---|
| <5.5 mmol/L(100 mg/dL) | SD<0.34mmol/L(6.0mg/dL) |
| 25.5 mmol/L(100mg/dL) | CV<6.0% |